You are here
ProCeas® H2 purity analyzer
The ProCeas® H2 purity is a complete pre-calibrated laser infrared spectrometer for detection of ppb-level impurities in hydrogen streams.
Key features:
- Patented OFCEAS® + LPS® technologies for ultra accurate, real-time measurements
- Can be combined with GC/FID-TCD for full compliance with ISO14687 – validated through MetroHyVe project
- Highest selectivity without cross-interference
- No zero gas or carrier gases required, reducing operational complexity
- Wide, dynamic measuring range
- Available in pressurized enclosure (ATEX, IECEx, cUL)
Key benefits:
- Long-lasting accuracy: On-line measurement, with no need for re-calibration or sensor replacement
- Reduced operational costs and complexity: Multi-point measurement, no zero gas or carrier gases required, easy to use with low maintenance
- Helps protecting infrastructure: Identifies contaminants in real-time enabling process adjustments that reduce fuel cells degradation and related maintenance costs
- Fast decision-making: Real-time, accurate data supporting process control and gas quality assurance
| ANALYTICAL SPECIFICATIONS | ||||
|---|---|---|---|---|
| Gas | Rangea | LODb | ||
| min | max | min | max | |
| H2S | 50 ppm | 10 % | 2 ppb | 100 ppm |
| CH4 | 50 ppm | 100 % | 1 ppb | 100 ppm |
| CO | 100 ppm | 100 % | 1 ppb | 100 ppm |
| CO2 | 50 ppm | 100 % | 2 ppb | 100 ppm |
| Zero drift | none | |||
| Other parameters | on request | |||
A adjustable range on request
B limit of detection 3 Sigma